What is the temperature range for Cidex Opa?

What is the temperature range for Cidex Opa?

59 – 86°F
CIDEX ® OPA Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.55% ortho -phthalaldehyde in an aqueous base containing buffers, chelating agents and a corrosion inhibitor. It is stable at 15 – 30°C (59 – 86°F) for two years.

How do you test for cidex?

How to test: Completely submerge indicating pad of the strip into solution. Hold for 1 second and remove. Do not shake strip. Remove excess solution by standing the strip upright on a paper towel.

How long is cidex Opa good for after opening?

The shelf life of an un- opened bottle of CIDEX OPA Solution is 2 years. The solu- tion requires NO activation. After opening the bottle, pour CIDEX OPA Solution into a CIDEX Solution Tray or appropriate container. If there is still solution left in the bottle, the bottle can be stored up to 75 days.

How often should cidex test strips be tested?

It is recommended that CIDEX OPA Solution be tested before each usage with the CIDEX OPA Solution test strips in order to guard against dilution, which may lower the ortho-phthalaldehyde level of the solution below its MEC of 0.3%. WARNING: Do not use CIDEX OPA Solution beyond its maximum 14 day use life.

What is the pH of cidex?

CIDEX® OPA Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.55% ortho-phthalaldehyde in an aqueous base containing buffers,chelating agents and a corrosion inhibitor.

How does cidex sterilization work?

OPA 0.55% is a high-level disinfectant introduced in 1999 as Cidex OPA solution (Advanced Sterilization Products). OPA solution achieves HLD in 12 minutes at room temperature, and 5 minutes when used in an automatic endoscope reprocessor at a temperature of 25°C (77°F).

What is the difference between cidex and cidex Opa?

CIDEX OPA is different from CIDEX Activated Dialdehyde Solution. CIDEX Activated Dialdehyde Solution uses glutaraldehyde as the disinfectant while CIDEX OPA uses ortho-phthalaldehyde. Bladder cancer patients are more sensitive to ortho-phthalaldehyde.

How do you use OPA MetriCide?

Submerge the MetriCide OPA Plus Solution Test Strip indicator pad into the solution for two seconds and then remove. Shake the strip to remove excess solution. Hold the strip horizontally and read the results at 60 seconds. A yellow background dye is included in the indicator of each test strip.

How do you use CIDEX OPA Solution?

Immerse clean, dry instruments completely in the CIDEX® OPA Solution. applicable, fill all lumens. Cover the CIDEX® Solution Sterilizing Disinfecting Tray with a secure lid. Soak instruments for 12 minutes at 20°C to achieve high-level disinfection.

What color is cidex?

Is Cidex OPA a high level disinfectant?

CIDEX OPA Solution is a high level disinfectant when used or reused, according to the Directions for Use, at or above its Minimum Effective Concentration (MEC) as determined by CIDEX OPA Solution Test Strips, with an immersion time of at least 12 minutes for a reuse period not to exceed 14 days.

What is the shelf life of Cidex OPA solution?

• CIDEX® OPA Solution must be discarded after 14 days even if CIDEX® OPA SolutionTest Strips indicate a concentration above the Minimum Effective Concentration (MEC). 6. Rinse Instruments Following disinfection, rinse instruments

How do I activate Cidex® OPA solution?

The solution requires NO activation. • ®After opening the bottle, pour CIDEX OPA Solution into a CIDEX®Sterilizing/Disinfecting Tray or appropriate container. If there is still solution remaining in the bottle, the bottle can be stored up to 75 days.

How do I sterilize my instruments with Cidex®OPA?

• Immerse clean, dry instruments completely in the CIDEX®OPA Solution. • Ensure all instruments are completely submersed in CIDEX®OPA Solution, and if applicable, fill all lumens. • ®Cover the CIDEX Solution Sterilizing Disinfecting Tray with a secure lid. Soak instruments for 12 minutes at 20°C to achieve high-level disinfection.